PromedeusLab KIM-1/TIM-1/HAVCR1 ELISA, Human Kit

PromedeusLab KIM-1/TIM-1/HAVCR1 ELISA, Human

PL1001
Regular price
$685.00
Sale price
$685.00
Regular price
$813.00
Unavailable
Unit price
per 
Shipping calculated at checkout.

Description

Enzyme Immunoassay kit for the quantitative determination Kidney Injury Molecule 1 (KIM-1) in human urine and serum samples. Quality controls are included with the kit. Developed and manufactured in Europe by PromedeusLab.

Summary

The urinary kidney injury molecule 1 (KIM-1) participates in renal tissue damage and repair and is proposed as a biomarker of early and subclinical acute kidney injury (AKI). KIM-1 is also elevated in the urine of a significant fraction of patients apparently recovered from an AKI. Besides its potential utility in the early and subclinical diagnosis of renal damage, research suggests a new application of urinary KIM-1 in the non-invasive follow-up of renal repair after AKI (1).

Other Names

Kidney injury molecule 1 (KIM-1, KIM1), T-cell immunoglobulin and mucin domain-containing protein 1 (TIMD-1, TIMD1), T-cell immunoglobulin mucin receptor 1 (TIM, TIM-1, TIM1), Hepatitis A virus cellular receptor 1 (HAVcr-1, HAVCR, HAVCR1), T-cell membrane protein 1, CD365, UniProtKB# Q96D42.

Principle of Method

The microtiter plate is coated with the antibody specifically binding the Kidney Injury Molecule 1. The human serum or plasma is incubated in the plate with the capture antibody.

The specimen is washed out and the specifically bound protein is incubated with HRP-labelled detection antibody. Unbound reagent is then washed out. Horseradish peroxidase (HRP) bound in the complex reacts with the chromogenic substrate (TMB) creating the blue colour. The reaction is stopped by addition of STOP solution (H2SO4).

The absorbance values are measured at 450 nm (optionally 450/630 nm) and are proportional to the concentration of KIM-1 in the specimen. The concentration of KIM-1 in unknown samples is determined from the calibration curve which is created by plotting the absorbance values against the standard concentration values

Sample Types

Urine, Serum

Typical Standard Curve

Assay Range: 15.625 pg/mL - 1000 pg/mL

Example of a typical standard curve when using the PromedeusLab KIM-1/TIM-1/HAVCR1 Human ELISA Kit.

Sensitivity

The limit of detection, defined as a concentration of human KIM-1 giving absorbance higher than absorbance of blank + 3 standard deviations, is better than 45 pg/mL of sample.

Precision

Intra-Assay

Test sample Mean (pg/mL) SD CV
1
519
8.8
1.7%
2
287
15.6
5.4%

 

Inter-Assay (Run-to-Run)

Test sample Mean (pg/mL) SD CV
1
284
16
5.7%
2
114
15
13.3%

Accuracy

Dilution Linearity:

Test sample Dilution
Measured Concentration (pg/mL)
Expected Concentration (pg/mL)
Yield
1

509
- -
1 2x 230
254
90%
1 4x 106
127
83%
1 8x 52
64
82%
2

448
-
-
2 2x 196
224
87%
2 4x 99
112
89%
2 8x 54
56
96%

Spiking Recovery:

Test sample Spike (ng/mL)
Measured Concentration (pg/mL)
Expected Concentration (pg/mL)
Yield
1
- 236
- -
1 360
663
596
111%
1 180
446
416
107%
1 90
341
326
105%

Storage

The kit must be stored at 2–8 °C. The opened components can be stored for one week at 2–8 °C.

Usage

This assay is for in vitro research use only and is not for use in diagnostic procedures.

Resources

Summary Resources

  • Cuesta C, Fuentes-Calvo I, Sancho-Martinez SM, Valentijn FA, Düwel A, Hidalgo-Thomas OA, Agüeros-Blanco C, Benito-Hernández A, Ramos-Barron MA, Gómez-Alamillo C, Arias M, Nguyen TQ, Goldschmeding R, Martínez-Salgado C, López-Hernández FJ. Urinary KIM-1 Correlates with the Subclinical Sequelae of Tubular Damage Persisting after the Apparent Functional Recovery from Intrinsic Acute Kidney Injury. Biomedicines. 2022 May 10;10(5):1106. doi: 10.3390/biomedicines10051106. PMID: 35625842; PMCID: PMC9139078.